STATISTICS SHEET FOR FOX800

Meets ASTM C 518 and ISO 8301 standards




Specimen size 762 mm square from 0 to 305 mm thick (30" square 0 to 12" thick)

  • Microprocessor based - stand alone operation (no PC needed)
  • Solid state, state of the art electronic heating/cooling
  • Electronic control and resolution of 0.6 mV
  • Temperature range : -15°C (5°F) to 80°C (176°F) on both plates - heat flow up or down
  • Temperature control and resolution of 0.01°C
  • Dual zone heating/cooling to prevent edge heat loss
  • Four inches insulation around test chamber
  • Instrument Dimensions: 38"W by 24"H by 38" D (952mmW by 601mmH by 952mmD)
  • Weight ~250lb.s (110kg)
  • Input power: 220V / 50/60Hz / user switchable at 2000VA
  • Thirty day money back guarantee
  • Warranty - one full year parts and labor

TESTING

  • Accuracy of better than 1%, 0.2% Repeatability, 0.5% Reproducibility
  • Ease of use enables initiation of test in less than a minute
  • Respond to prompts on four line alphanumeric LCD display
  • Testing times of less than 20 minutes per setpoint for 1" sample
  • Repeated testing of one setpoint or entry of up to nine setpoints
  • Plate spacing precision of 0.001"
  • Auto readout of plate spacing, or user set-able (compressible specimen)
  • Print out through standard RS 232 interface

OPTIONS

WinTherm Software (Included)

  • Allows connection of FOX instruments to IBM compatible computer thru RS 232 interface
  • Provides total control of instrument
  • Ability to change test parameters (temperatures, equilibrium criteria, PID), during test
  • Ability to run two FOX instruments simultaneously on one host
  • Capacity to run as many as 8 FOX instruments simultaneously
  • Ability to use computer for other tasks while testing if still in Windows