STATISTICS SHEET FOR FOX200 Vacuum 



Meets ASTM C 518 and ISO 8301 standards

Specimen size 203mm (8") square and 0 to 51mm (2")thick

  • The Vacuum level is 10-3atm
  • Microprocessor based -stand alone operation (no PC needed)
  • Solid state, state of the art electronic heating/cooling
  • Electronic control and resolution of 0.6 mV
  • Temperature range : -20°C (-4°F) to 95°C (203°F) on both plates - heat flow up or down
  • Temperature control and resolution of 0.01°C
  • Weight 35lb.s (16kg)
  • Input power: 115V or 220V / 50/60Hz / user switchable at 350VA
  • Cooling water at 18°C or less
  • Warranty - one full year parts and labor
TESTING
  • Accuracy of better than 1%, 0.2% Repeatability, 0.5% Reproducibility
  • Ease of use enables initiation of test in less than a minute
  • Respond to prompts on four line alphanumeric LCD display
  • Testing times of less than 10 minutes per setpoint for 1" sample
  • Repeated testing of one setpoint or entry of up to nine setpoints
  • Plate spacing precision of 0.001''
  • Auto readout of plate spacing, or user set-able (compressible specimen)
  • Print out results on standard RS232 interface

 


OPTIONS

WinTherm Software
  • Allows connection of FOX instruments to IBM compatible computer thru RS 232 interface
  • Provides total control of instrument
  • Ability to change test parameters (temperatures, equilibrium criteria, PID), during test
  • Ability to run two FOX instruments simultaneously on one host
  • Capacity to run as many as 8 FOX instruments simultaneously

 


Product Lines

© 1997-2010 LaserComp, Inc.