STATISTICS SHEET FOR FOX200LT


Meets ASTM C 518 and ISO 8301 standards

Specimen size 203mm (8") square and 0 to 30mm (1.182") thick

  • Temperature range : -175°C (-283°F) to 50°C (122°F)
  • Temperature control and resolution of 0.01°C
  • One calibration that covers the whole temperature range is provided based on the NIST 1450b Standard
  • Warranty - one full year parts and labor

TESTING

Testing will require connection to a LN2 (Liquid Nitrogen) dewar. In order to reduce the use of LN2 the instrument will control the flow rate by a controlled solenoid valve on the dewar. A small portion of the exhaust gas is routed inside the test chamber in order to provide a dry atmosphere. The Instrument is capable of completing nine entered setpoints covering the entire temperature range within a day.

The guage pressure of the LN2 tank is 22 psig (151 kPa, 1.517 bars). Itwill use ~120 liters for a typical full range test of ~10 hours duration. Tests should begin at the lowest temperature and end on the highest. This is critical to removing the problems instroduced by condensation. The Lowest mean temperature is -165°C.

OPTIONS

  • WinTherm Software
    • Allows connection of FOX instruments to IBM compatible computer thru RS232 interface
    • Provides total control of instrument
    • Ability to change test parameters (temperatures, equilibrium criteria, PID), during test
    • Ability to run two FOX instruments simultaneously on one host
    • Capacity to run as many as 8 FOX instruments simultaneously