STATISTICS SHEET FOR FOX200LT

Meets ASTM C 518 and ISO 8301 standards
Specimen size 203mm (8") square and 0 to 30mm (1.182") thick
-
Temperature range : -175°C (-283°F) to 50°C (122°F)
-
Temperature control and resolution of 0.01°C
-
One calibration that covers the whole temperature range is provided based on
the NIST 1450b Standard
-
Warranty - one full year parts and labor
TESTING
Testing will require connection to a LN2 (Liquid Nitrogen) dewar. In order to
reduce the use of LN2 the instrument will control the flow rate by a controlled
solenoid valve on the dewar. A small portion of the exhaust gas is routed
inside the test chamber in order to provide a dry atmosphere. The Instrument is
capable of completing nine entered setpoints covering the entire temperature
range within a day.
The guage pressure of the LN2 tank is 22 psig (151 kPa, 1.517 bars). Itwill use
~120 liters for a typical full range test of ~10 hours duration. Tests should
begin at the lowest temperature and end on the highest. This is critical to
removing the problems instroduced by condensation. The Lowest mean temperature
is -165°C.
OPTIONS
-
WinTherm Software
-
Allows connection of FOX instruments to IBM compatible computer thru RS232
interface
-
Provides total control of instrument
-
Ability to change test parameters (temperatures, equilibrium criteria, PID),
during test
-
Ability to run two FOX instruments simultaneously on one host
-
Capacity to run as many as 8 FOX instruments simultaneously